Abstract

Sensitive and high-throughput analysis of trace elements in volume-limited biological samples is highly desirable for clinical research and health risk assessments. However, the conventional pneumatic nebulization (PN) sample introduction is usually inefficient and not well-suited for this requirement. Herein, a novel high-efficiency (nearly 100% sample introduction efficiency) and low-sample-consumption introduction device was developed and successfully coupled with inductively coupled plasma quadrupole mass spectrometry (ICP-QMS). It consists of a micro-ultrasonic nebulization (MUN) component with an adjustable nebulization rate and a no-waste spray chamber designed based on fluid simulation. The proposed MUN-ICP-QMS could achieve sensitive analysis at a low sampling rate of 10 μL min-1 with an extremely low oxide ratio of 0.25% where the sensitivity is even higher comparing to PN (100 μL min-1). The characterization results indicate that the higher sensitivity of MUN is attributed to the smaller aerosol size, higher aerosol transmission efficiency, and improved ion extraction. In addition, it offers a fast washout (20 s) and reduced sample consumption (as low as 7 μL). The absolute LODs of the studied 26 elements by MUN-ICP-QMS are improved by 1-2 orders of magnitude compared with PN-ICP-QMS. The accuracy of the proposed method was validated by the analysis of human serum, urine, and food-related certified reference materials. Furthermore, preliminary results of serum samples from patients with mental illnesses demonstrated its potential in the field of metallomics.