Abstract
This paper presents a common-path electro-optic modulated circular polariscope capable of performing the sequential measurements of the principal axis and retardance of the linearly birefringent media with high accuracy by means of two phase-locked extractions, and with a high dynamic range obtained by the design of optical configuration and the derivation of an easy algorithm. There exist an absolute error of 0.10° on average in the principal axis measurement of 36 times within the range of 0°–180°, and a relative error of 1.10% in the retardance measurement of a λ/2-wave plate by this method. Consequently, the standard deviations for the principal axis angle and retardance measurements are determined to be 0.0033° and 0.086°, respectively, triple standard deviations to the average values of principal axis angle and retardance are just 0.10% and 0.14% with high repeatability, respectively, and corresponding dynamic ranges of 0°–180°.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.