Abstract

The cause and removal of streaking artifacts from catheters within radiochromic 3D dosimeters was investigated to allow the use of these dosimeters for high dose rate (HDR) brachytherapy applications. The OSC-TV iterative reconstruction algorithm’s ability to remove the streaking artifacts from one or more catheters was validated using external beam irradiations and leuco crystal violet (LCV) micelle gels. One and two catheter HDR plans were then delivered to LCV gels. Dose volumes from the gels were reconstructed with the OSC-TV algorithm, calibrated, and registered to treatment planning system dose volumes, and were shown to be artifact free and in agreement with the expected values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.