Abstract
A high density polyethylene (RCH 1000) femoral head prosthesis was inserted in 5 large dogs. Periodic clinical and roentgenographic evaluations, and histopathological examination of the replaced joints as well as biomechanical evaluation of the prosthetic joints were done at intervals from 3 weeks to 15 months. HDP is NOT suitable for movable weight-bearing prosthesis articulating with cartilage or bone. Wear of the cartilage and the polyethylene occurred within a few weeks of function and became extensive proportionally to the length of use. The accumulating wear particles of HDP caused a significant foreign body granulomatous reaction which resulted in erosion of bone adjacent to the acetabulum in one hip.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.