Abstract

BackgroundRepetitive atrial activation patterns (RAAPs) during atrial fibrillation (AF) may be associated with localized mechanisms that maintain AF. Current electro-anatomical mapping systems are unsuitable for analyzing RAAPs due to the trade-off between spatial coverage and electrode density in clinical catheters. This work proposes a technique to overcome this trade-off by constructing composite maps from spatially overlapping sequential recordings. MethodsHigh-density epicardial contact mapping was performed during open-chest surgery in goats (n=16, left and right atria) with 3 or 22 weeks of sustained AF (249-electrode array, electrode distance 2.4 mm). A dataset mimicking sequential recordings was generated by segmenting the grid into four spatially overlapping regions (each region 6.5 cm2, 48±10% overlap) without temporal overlap. RAAPs were detected in each region using recurrence plots of activation times. RAAPs in two different regions were joined in case of RAAP cross-recurrence between overlapping electrodes. We quantified the reconstruction success rate and quality of the composite maps. ResultsOf 1021 RAAPs found in the full mapping array (32±13 per recording), 328 spatiotemporally stable RAAPs were analyzed. 247 composite maps were generated (75% success) with a quality of 0.86±0.21 (Pearson correlation). Success was significantly affected by the RAAP area. Quality was weakly correlated with the number of repetitions of RAAPs (r=0.13, p<0.05) and not affected by the atrial side (left or right) or AF duration (3 or 22 weeks of AF). ConclusionsConstructing composite maps by combining spatially overlapping sequential recordings is feasible. Interpretation of these maps can play a central role in ablation planning.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call