Abstract
We have improved the optical characteristics of aluminum-coated fiber probes used in near-field scanning optical microscopy by milling with a focused ion beam. This treatment produces a flat-end face free of aluminum grains, containing a well-defined circularly-symmetric aperture with controllable diameter down to 20 nm. The polarization behavior of the tips is circularly symmetric with a polarization ratio exceeding 1:100. The improved imaging characteristics are demonstrated by measuring single molecule fluorescence. Count rates increase more than one order of magnitude over unmodified probes, and the molecule images map a spatial electric field distribution of the aperture in agreement with calculations.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.