Abstract
High radiocarbon sample throughput is an increasingly important requirement of accelerator mass spectrometry (AMS) systems, which in turn requires high source output. To accomplish this, HVE AMS systems use a model SO-110B ion source, which can run at outputs of more than 200 μA 12C. It is known that high source output may compromise precision. To quantify this effect, we have tested an HVE 1MV AMS system for the influence of medium (100 μA) and high (200 μA) source outputs on the precision. These early measurements indicate for a 200 μA 12C source output, equivalent to a ≃1 kHz 14C count rate for Ox II samples, a 13C/12C precision of 1% and a 14C/12C precision of 3% can be obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.