Abstract

The objective of this study was to evaluate the accuracy of a beta prototype version of a new portable blood glucose meter in feline patients. 60 client-owned cats. In this prospective study, 3-mL blood samples were collected from each cat and analyzed in triplicate using a beta prototype device (AlphaTRAK 3 [AT3]) and by a reference lab standard immediately after collection. Accuracy of the AT3 device was determined in accordance with the International Organization of Standardization (ISO) 15197:2013 criteria, including Bland-Altman plotting and consensus error grid analysis. A Passing-Bablok regression analysis was also performed. 96% of feline measurements fell within the ISO accuracy threshold, and 100% of measurements fell within zones A and B of the consensus error grid, meeting the ISO accuracy requirements. There was no significant bias in the data according to the Bland-Altman analysis. Within the full range of glucose concentrations (20 to 750 mg/dL) the correlation coefficient between the AT3 and the reference lab standard was 0.99. There was no significant constant or proportional bias present in the data. The AT3 device met the ISO requirements and is accurate for measurement of blood glucose concentrations in cats.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.