Abstract

We have developed a planar absolute radiometer for room temperature (PARRoT) that will replace the legacy C-series calorimeter as the free-space continuous-wave laser power detector standard at the National Institute of Standards and Technology (NIST). This instrument will lower the combined relative expanded measurement uncertainty (k = 2) from 0.84 % to 0.13 %. PARRoT's performance was validated by comparing its response against a transfer standard silicon trap detector traceable to NIST's primary standard laser optimized cryogenic radiometer (LOCR) and against the C-series calorimeter. On average, these comparisons agreed to better than 0.008 % and 0.05 %, respectively.