Abstract

AbstractThe vapor pressure of hexagonal (Ih) water ice was measured over the temperature range 175 K to 253.4 K and referenced to the value at the triple point of water. This experiment combined a highly accurate humidity generation system containing an ice‐coated saturator (millikelvin‐level temperature control) to provide humidified streams of nitrogen to a cavity‐enhanced laser absorption spectrometer. The measured ice vapor pressures had relative standard uncertainties ranging from 0.4% to 0.7% over the entire temperature range. We demonstrate that these measurements validate thermodynamic correlations for ice vapor pressure based on integration of the Clapeyron equation. Moreover, they also indicate that some commonly used vapor pressure correlations are inaccurate and should be avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.