Abstract
The Physikalisch-Technische Bundesanstalt determined the directional spectral emissivities of several widely used black coatings: Nextel 811-21, Herberts 1534, Aeroglaze Z306 and Acktar Fractal Black. These are and were often applied in different industrial and scientific applications. The measurements are taken angularly resolved over a range from 10{^{circ }} to 70{^{circ }}. They cover the temperature range typical for the application of the respective coating and a wide wavelength range from 4~upmu hbox {m} to 100~upmu hbox {m}. The respective directional total emissivities and hemispherical total emissivities are given as well. The measurements were taken under vacuum at the reduced background calibration facility to achieve low uncertainties and avoid atmospheric interferences. Additionally, some measurements were taken with the emissivity measurement setup in air.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.