Abstract

Miniature mass spectrometers are powerful tools for on-site chemical analysis in the fields of homeland security, personal healthcare, and environmental monitoring. This study presents a novel hexapole-assisted continuous atmospheric pressure interface for a high-pressure photoionization miniature ion trap mass spectrometer (HA-HPPI-IT). Efficient ion transmission was achieved by combining radial focusing by an RF electric field and axial driving by gas flow, which was demonstrated by SIMION simulation and experimental verification. The pressure in the ionization-transmission chamber and the inner diameter of the skimmer were optimized, which helped in determining the number density of product ions and affected the ion transmission in the hexapole, respectively. After systematic optimizations, about 16-fold increase in signal intensity was achieved as the RF amplitude was varied from 140 to 400 Vpp, and a limit of detection of 1 ppbv was obtained. In addition, the HA-HPPI-IT exhibited high stability and the relative standard deviation was as low as 5.47%. Finally, the apparatus was applied for discovering the simulated spot for illicit drug synthesis by detecting toluene and propiophenone released to air and monitoring the evolutions of perchloroethylene residues from dry-cleaned clothes.