Abstract

While modern immunoassays provide sensitive and specific means for the quantitation of cytokines in biological fluids, heterophile antibodies are still a well-recognized cause of interference in the measurement of cytokines in these assays. We have developed a multiplexed fluorescent microsphere immunoassay for the simultaneous quantification of 10 cytokines in only 75 microl of serum. During the development of this multiplexed assay, the amount of assay interference due to heterophile antibodies was also determined, and methods for detecting heterophile interference and minimizing its effect were incorporated into the assay. Heterophile antibodies resulted in significantly elevated cytokine values compared to those of normal blood bank samples. These falsely elevated values, and thus the components of the assay the heterophile antibodies were binding to, were identified through the use of internal controls. This information was then used to design assay-specific blockers and absorbents that were shown to significantly reduce falsely elevated cytokine values while not affecting the standard and control values. The fluorescent multiplexed microsphere-based immunoassay can be used to quantitate multiple cytokines from a single sample and should be a useful tool in furthering our understanding of the role of cytokines in disease processes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.