Abstract

Here the geometries that maximise detection efficiency of heterogeneous scintillators used to detect beta particles in aqueous solutions by maximising energy deposition are described. The determination of the geometry was achieved with the Monte Carlo code Geant4 using CaF2:Eu scintillator as a pertinent case study, and validated with experimental data using single crystal CaF2:Eu and heterogeneous CaF2:Eu scintillators. Both 2D and 3D structures composed of arrays of primitive unit cells of packed spheres were examined to find the optimal geometry to maximise detection of volumetric sources of tritium and aqueous Carbon 14 and Lead 210. The 2D structures were evaluated relative to a single crystal scintillator and results show the detection efficiency of the 2D structures is maximised when the sphere radius is c.a. 0.46x the maximum track length of the beta particle in the scintillator. Data for the 3D structures show that the efficiency is maximised when the sphere radius is minimised, but it is further shown that practical issues limit the minimum radius that can be used for transient radiological contamination monitoring.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.