Abstract

Background The study aimed to evaluate a fully automated chemiluminescent immunoassay and compared it with a quantitative RNA assay and anti-HCV assay to verify the utility of this automated Ag assay as an alternative method for hepatitis C diagnosis. Methods A total of 229 serum samples previously tested for anti-HCV concentrations by the Architect Anti-HCV assay, were selected for HCV RNA testing by real time RT-PCR kit (Shanghai ZJ Bio-Tec Co., Ltd) and 125 specimens were tested for HCV Ag by the Architect HCV core Antigen kit. Results The log10 HCVAg and HCV RNA concentrations were highly correlated [ r = 0.834); with HCV RNA as the comparator test, HCVAg had 100% specificity, 100% positive predictive value (PPV) and 94.8% sensitivity. We found 1 pg/mL of total HCV core Ag is equivalent to approximately 6607HCV RNA international units (IU)/mL. Receiver operator characteristic curve analysis showed that the area under the curve of HCV core Ag (0.989) was greater than HCV Ab (0.871). HCV Ag concentrations and RNA-to-Ag ratio of the groups for HCV RNA concentrations ≤105 and >105 IU/mL were both significantly different from each other ( P < 0.05). Conclusion The Architect HCV core Ag assay may be an alternative method for hepatitis C diagnosis, performed on the same analytical platform and sample as the anti-HCV assay, shortening the diagnostic window period, demonstrating good correlation with HCV RNA assay with high specificity and positive predictive value.