Abstract

A magnetically suspended centrifugal pump (MSCP), developed for long-term ventricular assist, is reliable and durable because it has no shaft or seal. However, with nonvalve pumps such as a MSCP, regurgitation occurs when they accidentally stop without cannula clamping. We investigated the hemodynamics during temporary stoppage of a MSCP being used as a left ventricular assist system (LVAS), comparing two inflow cannulation sites. In four sheep (weight, 35-45 kg), microspheres were injected into the left main coronary artery to induce heart failure. An outflow cannula was sutured onto the descending aorta, and two inflow cannulae were inserted into the left atrium and the left ventricle. The MSCP was stopped with both the left ventricular cannula and left atrial cannula clamped, and the hemodynamics and P-V loops were recorded. Each cannula was then unclamped in order, and similar parameters were recorded. LVEDP increased at unclamping of the left ventricular cannula (ULVC), and rose further at unclamping of the left atrial cannula (ULAC). Aortic pressure did not change at ULVC, but decreased at ULAC. The effective systemic flow that subtracted the regurgitant flow through the MSCP from left ventricular output was half at ULVC and almost 0 at ULAC. When stopping centrifugal pumps without circuit clamping, hemodynamic deterioration is less at ULVC than at ULAC. This finding suggests that left ventricular inflow cannulation is recommended to allow more time in emergency situations.