Abstract

Renal replacement therapy by hemodialysis requires a permanent vascular access. Implantable ports offer a potential alternative to standard vascular access strategies although their development is limited both in number and extent. We explored the fluid dynamics within two new percutaneous bone-anchored dialysis port prototypes, both by in vitro experiments and computer simulation. The new port is to be fixed to bone and allows the connection of a dialysis machine to a central venous catheter via a built-in valve. We found that the pressure drop induced by the two ports was between 20 and 50 mmHg at 500 ml/min, which is comparable with commercial catheter connectors (15-80 mmHg). We observed the formation of vortices in both geometries, and a shear rate in the physiological range (<10,000s-1), which is lower than maximal shear rates reported in commercial catheters (up to 13,000s-1). A difference in surface shear rate of 15% between the two ports was obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.