Abstract

Electrical cardiometry (EC) is an impedance-based monitor that provides noninvasive, real-time hemodynamic assessment. However, the reference values for neonates have not been established. EC (Aesculon) was applied to hemodynamically stable preterm and term infants. Hemodynamic variables included cardiac output (CO), cardiac index (CI), stroke volume (SV) and heart rate (HR). Their gestational age (GA), weight and body surface area (BSA) were recorded. A total of 280 neonates were studied. Their GA ranged from 26(5/7) to 41(4/7) weeks, weight 800 to 4420 g and BSA 0.07 to 0.26 m(2). CO was positively correlated to GA, weight and BSA (r=0.681, 0.822, 0.830, respectively; all P<0.001). Using regression analysis, CO was most significantly correlated to BSA. Mean CI was 2.55±0.37 l min(-1) per m(2). Hemodynamic reference by EC is notably distinct among neonates of diverse maturity. CO is most closely correlated to BSA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.