Abstract

Background To test whether later intra-aortic balloon pump (IABP) deflation approaching or simultaneous with left ventricular ejection would improve hemodynamics and myocardial efficiency with the use of new balloon deflation methods, 4 IABP timing techniques were evaluated in 43 patients. Methods and Results Later balloon deflation produced significantly greater percentage changes in mean aortic pressure (6% vs 1%), systolic pressure time index (–27% vs –20%), diastolic pressure time index (35% vs 19%), and the systolic pressure-time index/diastolic pressure-time index ratio (97% vs 51%), respectively. However, these changes increased peak systolic pressure (–15% vs –11%). Cardiac output and stroke volume indexes were not significantly altered over the 4 settings. Conclusions These data suggest that systemic hemodynamics and myocardial efficiency may be improved by later balloon deflation approaching left ventricular ejection in comparison to conventional IABP timing. (Am Heart J 1999;137:1129-6.)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.