Abstract

Background/Aims: Anticoagulation in hemodialysis (HD) patients at high risk of bleeding remains an intractable problem. Simulating endothelial cells by releasing anticoagulant on the membrane may be a promising alternative. Methods: We modified a polyacrylonitrile (PAN) dialyzer by loading argatroban into its membrane and verified its anticoagulation efficiency and its influence on coagulation markers such as activated partial thromboplastin time (aPTT), D-dimer and thrombin-antithrombin complex (TAT) in an animal HD model. Results: All HD sessions with the argatroban dialyzer were completed successfully with either no or minimal fiber clotting. D-dimer was lower in the argatroban group than in the PAN group; TAT and aPTT values were similar in the two groups, which suggests better anticoagulation and a similar influence on the coagulation system. Conclusion: HD with the argatroban dialyzer is feasible, safe and simple, and could be used in patients at high risk of bleeding.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.