Abstract

The authors retrospectively reviewed the use of ultrahigh-pressure angioplasty balloons at atmospheric pressures at or above the manufacturer recommended burst pressure (30 atm) for the treatment of resistant hemodialysis-related venous stenosis at their institution. In seven of 87 procedures, high-pressure angioplasty (up to 27 atm) was unsuccessful. By coupling new balloon technology with aggressive inflation pressures, 100% technical success was achieved in the treatment of stenoses that were resistant to high-pressure angioplasty in these seven procedures. This approach could potentially offer cost savings compared with the costs of other previously described treatment methods for resistant lesions, such as atherectomy devices and cutting balloons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.