Abstract

A novel microsampling device, Mitra(TM), was evaluated for bioanalysis of E6005 and its O-desmethylated metabolite in human whole blood using an UPLC-MS. A constant volume of blood samples was absorbed onto the tip of Mitra, the analytes were extracted by various solvents and then detected by UPLC-MS. Recovery of the analytes was high in acetonitrile-water (1:1, v/v) but was dependent on hematocrit (Hct) without sonication process, which led to biased accuracy at low and high Hcts. Inclusion of sonication process in extraction improved recovery at high Hct to yield acceptable accuracy across Hcts. Optimization of extraction process to achieve high recovery regardless of Hct is critical in accurate bioanalysis via Mitra.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.