Abstract

This paper presents a new permeation measurement facility based on the difference method, which is shown to be capable of measuring gas fluxes through different polymeric membranes. In order to correct for the background pressure increase in the measurement chamber, a reference chamber has been adapted and connected to it. The permeation flux was examined by measuring the pressure difference between the reference and measurement chambers. Pure helium permeation was performed through polydimethylsiloxane (PDMS) and polyethylene terephthalate (PET) membranes using this method. The uncertainty of the permeability was evaluated. The results show that the introduction of reference chamber can mitigate the influence of the residual pressure in the measurement chamber. This new method allows simplified and visualized measurement of the permeation fluxes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.