Abstract

A three-part analysis was undertaken to increase understanding of the occurrence of pressure ulcers in lithotomy positions. An innovative measuring device was used to determine capillary pressure. Ankle blood pressure was measured compared to ankle height in 11 participants. Ankle systolic and diastolic pressure decreased approximately 20 mmHg per foot of elevation. Calf and heel capillary-support pressures were measured in 15 participants in the standard lithotomy position. Capillary-support pressure for the calf was substantially less than for the heel. Heel capillary-support pressures were measured in 16 participants in the high lithotomy position. As heel height increased, capillary-support pressure also increased.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.