Abstract

There are different types of transurethral prostatic surgeries and the complication profiles are different. This study aims to compare the heat damage zones (HDZ) created by five different technologies in a pig liver model. Monopolar resection, bipolar resection, electrovaporization, and Greenlight™ lasers of 120 and 180 W were used to remove fresh pig liver tissue in a simulated model. Each procedure was repeated in five specimens. Two blocks were selected from each specimen to measure the three deepest HDZ. The mean of HDZ was 295, 234, 192, 673, and 567 μm, respectively, for monopolar resection, bipolar resection, electrovaporization, Greenlight laser 120 W, and Greenlight laser 180 W, respectively. The Greenlight laser produced one to three times deeper HDZ than the other energy sources (p=0.000). Both 120 and 180 W Greenlight lasers produced deeper HDZ than the other energy sources. Urologists need to be aware of HDZ that cause tissue damage outside the operative field.