Abstract
Insensible water loss, oxygen consumption, and carbon dioxide production were measured in eight premature infants under four different conditions: in conventional single-walled incubator with and without plastic heat shield, and under radiant warmer with and without heat shield. IWL was greater under the radiant warmer (3.40 +/- 1.50 ml/kg/hour, mean +/- SD) than in the incubator (2.37 +/- 1.15 ml/kg/hour) when both were compared without heat shield. Addition of the heat shield reduced IWL in the incubator (2.13 +/- 0.76 ml/kg/hour) but not under the radiant warmer (3.37 +/- 0.94 ml/kg/hour). There were no significant differences in VO2 or respiratory quotient between any two of the four study conditions.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.