Abstract

This paper proposes an algorithm for eliminating motion artifacts in heart rate monitoring using a photoplethysmography (PPG) sensor. No additional hardware, such as accelerometers or other reference signals, is needed. The algorithm improves the quality of PPG signals by implementing a comb filter during signal processing. The proposed slow–fast adaptive comb filter (SFACF) distinguishes motion artifacts in PPG signals based on harmonic and continuous features of the heart rate. The effectiveness of fundamental and harmonic frequency enhancement in real time is experimentally demonstrated, with the heart rate updated every second. The performance of the SFACF, which does not require additional reference signals, is compared with those of adaptive noise cancellation (ANC) and spectral peak search-comb filtering (SPS-CF). It is found that for a low-quality PPG signal, the SFACF outperforms ANC and SPS-CF. Finally, measurements obtained using a commercial heart rate monitor and the proposed algorithm are shown to be highly correlated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.