Abstract
A new and in part automated headspace solid-phase microextraction method for quantitative determination of the highly toxic rodenticide fluoroacetic acid (FAA) in serum and other biological samples has been developed. FAA and deuterated acetic acid (internal standard) were extracted from acidified samples by a StableFlex divinylbenzene–Carboxen on polydimethylsiloxane fibre. The acids were derivatised on the fibre in-situ with 1-pyrenyldiazomethane and detected using gas chromatography–mass spectrometry with electron impact ionisation and selected ion monitoring. The calibration curve for FAA in serum was linear over the range from 0.02 to 5 μg/ml, with limits of detection and quantification of 0.02 and 0.07 μg/ml, respectively. The method was also tested with spiked whole blood, urine, stomach contents and kidney samples. It was sufficiently reliable, reproducible and sensitive for use in routine forensic toxicology applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.