Abstract
A new MIMS-derived technique, headspace membrane introduction mass spectrometry (HS-MIMS), is described for direct trace level analysis of volatile organic compounds (VOCs) in soil and other dry or wet solid matrixes. A silicone membrane interface is placed about 15 cm from the ion source, and a closed airspace (headspace) is created by connecting a toggle valve to the 1/4 in. tubing that connects the membrane interface to the ion source. For the VOC analysis, the headspace is evacuated and the solid sample vessel is heated to 90 degrees C. The VOCs are rapidly desorbed from the sample, pervaporated through the membrane, and preconcentrated for 4 min in the evacuated headspace. Then, the toggle valve is opened and the trapped VOCs are released into the ion source region of a quadrupole mass spectrometer. By electron ionization and selected-ion monitoring, a relatively sharp and intense peak is obtained and used for quantification. The HS-MIMS analysis shows excellent linearity and reproducibility and detection limits for many VOCs typically of 50-100 ng/kg (ppt).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.