Abstract
A headspace gas chromatographic (GC) method was developed to determine 1,3-butadiene (1,3-BD) in simulated saliva in contact with chewing gum. The calibration graph was linear, and the limit of detection was 0.004 mg/L, which is well below the migration limit for this substance. The headspace GC method provides rapid and reliable analysis for monitoring 1,3-BD migration from chewing gum into simulated saliva. In this paper, we report headspace methodology for sensitive determination of 1,3-BD in chewing gum and results of selected analyses, enabling preliminary assessment of possible exposure to 1,3-BD through migration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.