Abstract

We propose an active noise control (ANC) system for reducing periodic noise generated in a high magnetic field such as noise generated from magnetic resonance imaging (MRI) devices (MRI noise). The proposed ANC system utilizes optical microphones and piezoelectric loudspeakers and consists of a head-mounted structure to control noise near the user's ears. Moreover, internal model control (IMC)-based feedback ANC is employed because the MRI noise includes some periodic components and is predictable. Our experimental results demonstrate that the proposed ANC system (head-mounted structure) can significantly reduce MRI noise by approximately 30 dB in a high field in an actual MRI room even if the imaging mode changes frequently.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.