Abstract
Optoacoustic tomography has been established as a powerful modality for preclinical imaging. However, efficient whole-body imaging coverage has not been achieved owing to the arduous requirement for continuous acoustic coupling around the animal. In this work, we introduce panoramic (3600) head-to-tail 3D imaging of mice with spiral volumetric optoacoustic tomography (SVOT). The system combines multi-beam illumination and a dedicated head holder enabling uninterrupted acoustic coupling for whole-body scans. Image fidelity is optimized with self-gated respiratory motion rejection and dual speed-of-sound reconstruction algorithms to attain spatial resolution down to 90 µm. The developed system is thus highly suitable for visualizing rapid biodynamics across scales, such as hemodynamic changes in individual organs, responses to treatments and stimuli, perfusion, total body accumulation, or clearance of molecular agents and drugs with unmatched contrast, spatial and temporal resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.