Abstract

Head column-field amplified sample stacking (HC-FASS) is one of the most powerful concentration techniques in capillary electrophoresis. In the present work we demonstrate that a laboratory-designed pneumatic sampler not only enables us to easily perform HC-FASS from stagnant sample solutions but also permits us to carry out HC-FASS from sample stream, which creates an opportunity to lower the detection limit of analytes further. The influence of stacking time in case of the flowing sample on peak efficiencies and resolution is discussed. It is demonstrated that the sample plug length can be kept constant by monitoring the current in the capillary, which controls a feedback system based on backpressure at the capillary outlet.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.