Abstract
A 69-year-old male diabetic patient had an abnormally low HbA1C value of 2.8%, which was inconsistent with his elevated fasting plasma glucose of 8.2 mmol/l. Hb analysis disclosed that the abnormal Hb was Hb Ube-2 [alpha68 (E17) Asn --> Asp] and it accounted for 21.5% of the total Hb. Since the glycated abnormal Hb emerged at the same position as did HbF on high performance liquid chromatography, the HbA1C value was falsely low. The present case demonstrates that Hb Ube-2 is one of the abnormal Hbs in which caution should be exercised when monitoring diabetic control.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.