Abstract

In the first two documents [Ph. Hubert, J.J. Nguyen-Huu, B. Boulanger, E. Chapuzet, P. Chiap, N. Cohen, P.A. Compagnon, W. Dewé, M. Feinberg, M. Lallier, M. Laurentie, N. Mercier, G. Muzard, C. Nivet, L. Valat, J. Pharm. Biomed. Anal. 36 (2004) 579–586; Ph. Hubert, J.J. Nguyen-Huu, B. Boulanger, E. Chapuzet, P. Chiap, N. Cohen, P.A. Compagnon, W. Dewé, M. Feinberg, M. Lallier, M. Laurentie, N. Mercier, G. Muzard, C. Nivet, L. Valat, E. Rozet, J. Pharm. Biomed. Anal., in press], a recent SFSTP Commission on the validation of analytical procedure has introduced a harmonized approach for the validation of analytical procedures. In order to complete this guide, the statistical methodology allowing to correctly conclude about the validity of a procedure is proposed in this third part of the guide. Indeed all the steps to obtain the decision tool namely the accuracy profile are described and illustrated step by step by a numerical example. This tool, based on the concept of total error (bias+standard deviation) build with a β-expectation tolerance interval, allows to easily take the right decision and simultaneously minimizing the risk of the future use of the analytical procedure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.