Abstract
BackgroundMonopolar cautery is the most commonly used surgical cutting and hemostatic tool for head and neck surgery. There are newer technologies that are being utilized with the goal of precise cutting, decreasing blood loss, reducing thermal damage, and allowing faster wound healing. Our study compares thermal damage caused by Harmonic scalpel and CO2 laser to cadaveric tongue.MethodsTwo fresh human cadaver heads were enrolled for the study. Oral tongue was exposed and incisions were made in the tongue akin to a tongue tumor resection using the harmonic scalpel and flexible C02 laser fiber at various settings recommended for surgery. The margins of resection were sampled, labeled, and sent for pathological analysis to assess depth of thermal damage calculated in millimeters. The pathologist was blinded to the surgical tool used. Control tongue tissue was also sent for comparison as a baseline for comparison.ResultsThree tongue samples were studied to assess depth of thermal damage by harmonic scalpel. The mean depth of thermal damage was 0.69 (range, 0.51 - 0.82). Five tongue samples were studied to assess depth of thermal damage by CO2 laser. The mean depth of thermal damage was 0.3 (range, 0.22 to 0.43). As expected, control samples showed 0 mm of thermal damage. There was a statistically significant difference between the depth of thermal injury to tongue resection margins by harmonic scalpel as compared to CO2 laser, (p = 0.003).ConclusionIn a cadaveric model, flexible CO2 laser fiber causes less depth of thermal damage when compared with harmonic scalpel at settings utilized in our study. However, the relevance of this information in terms of wound healing, hemostasis, safety, cost-effectiveness, and surgical outcomes needs to be further studied in clinical settings.
Highlights
Monopolar cautery is the most commonly used surgical cutting and hemostatic tool for head and neck surgery
Three cadaveric tongue samples were analyzed for thermal damage with the harmonic scalpel
Five cadaveric tongue samples were analyzed for thermal damage with the photonic band gap fiber assembly (PBFA) carbon dioxide laser
Summary
Monopolar cautery is the most commonly used surgical cutting and hemostatic tool for head and neck surgery. Our study compares thermal damage caused by Harmonic scalpel and CO2 laser to cadaveric tongue. Thermal damage can have deleterious effects on wound healing, safety and clinical outcomes. Alternative technologies such as the harmonic scalpel (Figure 1) and carbon dioxide (CO2) laser are gaining direct visualization is limited. This new technology has added versatility to the use of the laser and is being employed in all areas of otolaryngology with good surgical results. Our objective was to compare the tissue effects of the harmonic scalpel and PBFA carbon dioxide laser in tongue resections using a human cadaveric model
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.