Abstract

Hardware-related complications have been commonly described after deep brain stimulation. We searched the PubMed database using the key words ‘adverse effects’ and ‘deep brain stimulation’. Out of 254 studies, we identified 10 articles that primarily addressed morbidity and hardware-related adverse effects in 922 patients. The most commonly reported hardware-related complications were infections (6.1% of the patients), migration or misplacement of the leads (5.1% of the patients), lead fractures (5.0% of the patients), and skin erosion (1.3% of the patients). Refinements in surgical technique, technological improvements, and a greater experience with the procedures will likely decrease the incidence of hardware-related side effects in the future.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.