Abstract

Access to and extensive use of fluorometric analyses is limited, despite its extensive utility in environmental transport and fate. Wide-spread application of fluorescent tracers has been limited by the prohibitive costs of research-grade equipment and logistical constraints of sampling, due to the need for high spatial resolutions and access to remote locations over long timescales. Recently, low-cost alternatives to research-grade equipment have been found to produce comparable data at a small fraction of the price for commercial equipment. Here, we prototyped and benchmarked performance of a variety of fluorometer components against commercial units, including performance as a function of tracer concentration, turbidity, and temperature, all of which are known to impact fluorometer performance. While component performance was found to be comparable to the commercial units tested, the best configuration tested obtained a functional resolution of 0.1 ppb, a working concentration range of 0.1 to >300 ppb, and a cost of USD 59.13.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.