Abstract

The prevalence rate of neonatal jaundice can reach 80%, of which 5% may develop dangerous hemolytic jaundice. The blood test for obtaining bilirubin and hemoglobin concentration is the gold standard for diagnosing hemolytic jaundice; however, frequently drawing blood from jaundice neonates for the screening purpose is not practical. We have developed a handheld diffuse reflectance spectroscopy system to noninvasively determine the bilirubin and hemoglobin levels in neonates. Our study showed that the correlation coefficients were 0.95 and 0.80 for bilirubin and hemoglobin between the results from the blood tests and our handheld system, respectively. This handheld system could be an effective tool for screening hemolytic jaundice.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.