Abstract

Translation of photoacoustic imaging into the clinic is a major challenge. Handheld real-time clinical photoacoustic imaging systems are very rare. Here, we report a combined photoacoustic and clinical ultrasound imaging system by integrating an ultrasound probe with light delivery for small animal imaging. We demonstrate this by showing sentinel lymph node imaging in small animals along with minimally invasive real-time needle guidance. A clinical ultrasound platform with access to raw channel data allows the integration of photoacoustic imaging leading to a handheld real-time clinical photoacoustic imaging system. Methylene blue was used for sentinel lymph node imaging at 675 nm wavelength. Additionally, needle guidance with dual modal ultrasound and photoacoustic imaging was shown using the imaging system. Depth imaging of up to 1.5 cm was demonstrated with a 10 Hz laser at a photoacoustic imaging frame rate of 5 frames per second.