Abstract

This paper describes a measurement method developed at National Institute of metrological Research (INRIM) to calibrate picoammeters in dc current from 100 pA to 100 nA. The current source is based on a traceable to the dc resistance national standard 10 × 100 MΩ Hamon resistor developed at INRIM and on a traceable to the dc voltage national standard high precision dc voltage calibrator. The expanded uncertainties of the method for the calibration of picoammeters span from 9.4 × 10 −4 for the gain of a picoammeter at 100 pA to 4.0 × 10 −4 for the gain at 100 nA. A detailed uncertainties budget at 10 nA level and the results of a comparison with a different technique are also reported.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.