Abstract

If the anaesthetic circle system is arranged to increase the humidity of fresh anaesthetic gases by placing the carbon dioxide absorbent canister between the fresh gas inlet and the patient, drying of the soda-lime can occur. Very dry soda-lime adsorbs significant quantities of halothane. Using fresh soda-lime, effluent halothane concentration reached 50% of the input concentration in 35s, but this time increased to 500 s when dry soda-lime was used. The use of dry soda-lime can result in a slow inhalation induction or in the release of absorbed halothane during a subsequent anaesthetic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.