Abstract
The calibration of several ions (Cl −, Br −, F − and OH −) measured with an ion selective electrodes (ISE) array has been carried out in the presence of interferents using an experimental design and multivariate calibration methods. Partial least squares regression and principal component regression do not seem to improve the test set prediction compared to multivariate linear regression. In the case of very slight or no interference on the ISE, each ion can be determined using the corresponding ISE and univariate calibration methods, but the use of multivariate methods does not lead to worse results.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.