Abstract
Half-life measurements have been carried out at LNHB for 88Y using a 4π γ-ionisation chamber and for 89Sr using a proportional counter. The determined half-life values and associated standard uncertainties are 106.63±0.05 d for 88Y and 50.65±0.05 d for 89Sr, being consistent with relevant values reported in literature. Based on the present results and relevant literature values revised recommended half-life values and associated standard uncertainties are proposed, viz. 106.626±0.021 d for 88Y and 50.57±0.03 d for 89Sr.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.