Abstract
Solid-state detectors (SSDs) may be used along with a lead collimator for half-value layer (HVL) measurement using computed tomography (CT) with or without a tin filter. We aimed to compare HVL measurements obtained using three SSDs (AGMS-DM+ , X2 R/F sensor, and Black Piranha) with those obtained using the single-rotation technique with lead apertures (SRTLA). HVL measurements were performed using spiral CT at tube voltages of 70-140kV without a tin filter and 100-140kV (Sn 100-140kV) with a tin filter in increments of 10kV. For SRTLA, a 0.6-cc ionization chamber was suspended at the isocenter to measure the free-in-air kerma rate ( ) values. Five apertures were made on the gantry cover using lead sheets, and four aluminum plates were placed on these apertures. HVLs in SRTLA were obtained from decline curves. Subsequently, SSDs inserted into the lead collimator were placed on the gantry cover and used to measure HVLs. Maximum HVL differences of AGMS-DM+ , X2 R/F sensor, and Black Piranha with respect to SRTLA without/with a tin filter were - 0.09/0.6 (only two Sn 100-110kV) mm, - 0.50/ - 0.6mm, and - 0.17/(no data available) mm, respectively. These values were within the specification limit. SSDs inserted into the lead collimator could be used to measure HVL using spiral CT without a tin filter. HVLs could be measured with a tin filter using only the X2 R/F sensor, and further improvement of its calibration accuracy with respect to other SSDs is warranted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.