Abstract
A plate dialyser requiring an extracorporeal blood volume of 1.5 ml was developed to dialyse conscious rats. In experiments in vitro and in vivo its function was tested. The in vitro clearances of urea, creatinine and potassium were 126+/-9 ml/min/m2, 70.5+/-9 ml/min/m2, and 132.5+/-13 ml/min/m2, respectively. The method appears to be suitable for pharmacological and toxicological studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Research in experimental medicine. Zeitschrift fur die gesamte experimentelle Medizin einschliesslich experimenteller Chirurgie
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.