Abstract
Abstract Objective: For commissioning of source path mo-dels for ring tubes of Nucletron (now Elekta) CT/MR ring applicators, a method was established using a source position simulator (SPS) and a CT scanner. Methods: An acrylic glass support for ring tubes was constructed serving as defined coordinate system. With SPS, 40 positions were set in the ring in steps of 2.5mm from proximal to distal direction. In each position a CT-series was taken with ring plane inside scan plane using a GE Optima 580 scanner (slice width 0.625mm). Evaluation was done in multiplanar recon-struction mode under display settings revealing dummy source shape. With a 3D-cursor dwell position coordinate (center of Ir-192 core) was determined. Measurements were validated by film autoradiography in selected positions. Four rings were investigated: 60° Φ26, 45° Φ26, 60° Φ30, and 45° Φ30 with figures denoting angle between ring plane and stem and ring diameter in mm. Results: The method is feasible to reconstruct source dwell positions with an overall accuracy of ±0.3mm. Detected coordinates are consistent and in excel-lent agreement with those obtained by autoradiography. Comparison with Elekta’s source path models, however, re-vealed mean deviations of up to 3.5mm. Discussion: It is advisable to measure individual source paths and have them added to the data base of the treatment planning system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.