Abstract

Semiconductor diodes offer many advantages for clinical dosimetry: high sensitivity, real-time read-out, simple instrumentation, robustness and air pressure independence. The feasibility and usefulness of in vivo dosimetry with diodes has been shown by numerous publications, but very few, if any, refer to the utilization of diodes in electron beam dosimetry. The purpose of this paper is to present our methods for implementing an effective IVD program for external beam therapy with photons and electrons and to evaluate a new type of diodes. Methods of deciding on reasonable action levels along with calibration procedures, established according to the type of measurements intended to be performed and the action limits, are discussed. Correction factors to account for nonreference clinical conditions for new types of diodes (designed for photon and electron beams) are presented and compared with those required by older models commercially available. The possibilities and limitations of each type of diode are presented, emphasizing the importance of using the appropriate diode for each task and energy range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.