Abstract

The Dutch working group on 'Quality Criteria for Equipment Used in Diagnostic Radiology' has formulated guidelines providing technical criteria for equipment used in conventional diagnostic radiology. These guidelines are applicable to the technical parameters having a major impact on image quality and patient dose and include methods for testing. The following parameters are included: tube voltage, automatic exposure control, film processing, film-screen combination, light tightness and illumination of the dark room, half-value layer and filtration, light field, grid, focal spot size, viewing boxes and geometrical indicators. Each guideline consists of the following chapters: (1) Scope and field of application, (2) Background information, (3) Test procedure, (4) Test frequency, (5) Registration of observations, (6) Evaluation and interpretation, (7) Test report. Chapter 3 includes both the principles of the test method and a step by step description of the procedures. The principles of the test procedure provide a basis for adaptation to local circumstances. The step by step test procedure allows a quality control measurement to be performed with limited physical knowledge of the equipment. Chapter 6 includes limiting values. Draft guidelines were evaluated in practice in 20 hospitals. The final document has been accepted by the professional societies in the Netherlands and the Dutch Minister of Health as a reference set of tools to perform Quality Control of equipment used for conventional diagnostic radiology.