Abstract
An 8-channel AFE with a group-chopping instrumentation amplifier (GCIA) is proposed for bio-potential recording applications. The group-chopping technique cascades chopper switches to progressively swap channels and dynamically removes gain mismatch among all channels. An 8-phase non-overlapping clocking scheme is developed and achieves excellent between-channel gain mismatch characteristics. The dynamic offsets among all channels are mitigated by the GCIA as well. The GCIA is the first work that minimizes the gain mismatch across more than two channels. With the help of the group-chopping, combined with an area-efficient open-loop structure, the GCIA shows <0.04% between-channel gain mismatch, the lowest mismatch reported to date. The chip is fabricated in 0.18µm 1P6M CMOS, occupies only 0.017 mm2/Ch., consumes 2.1 μW/Ch. under 0.5 V supply and achieves an NEF of 2.1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Biomedical Circuits and Systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.